Sensitivity

Systematic Method Development and Troubleshooting

October 15, 2022
Matrix Effects in LC-MS/MS
Method Development, Sensitivity, Assessment of Matrix Effects, Trifluoroacetic Acid, TFA, Buffer, Ion Pairs

 Tips: Frequently, resolving the problem requires a combination of individual experimental solutions and systematic and logical investigation.

 Part 1 highlights rational method development and validation and includes troubleshooting tips for sound sample preparation, for analyte stability in biological matrices, and for chromatographic conditions and injection solvents.

 Part 2 further discusses development strategies for LC/MS/MS analysis of biopharmaceutical products including proteins and peptides. It will present troubleshooting tips for carryover, recovery and matrix effects, and selectivity; and for automation where possible to improve the overall reliability and efficiency of LC/MS/MS.

...

APCI General Review

March 17, 2022
Analysis of Pesticides in Food and Environmental Samples, edited by Jose L. Tadeo
Sensitivity, Modifier

The LC elute passing through a nebulizing needle, creates a fine spray. The spray passes through a heated ceramic tube, where the droplets are fully vaporized. The resulting gas/vapor mixture is then passed over a corona discharge needle, where the solvent vapor is ionized to create reagent gas ions.

For successful APCI, the analyte must be volatile and thermally stable and the mobile phase must be suitable for gas-phase acid–base reactions.

To vaporize the solvent and analyte, the APCI source is typically operated at vaporizer temperatures of 400 to 500 0C.

...

ESI General Review

March 17, 2022
Analysis of Pesticides in Food and Environmental Samples, edited by Jose L. Tadeo
Sensitivity, the ion evaporation mechanism, the charge residue mechanism

The LC eluate is sprayed into a spray chamber at atmospheric pressure in the presence of a strong electrostatic field and heated drying gas.

The electrostatic field occurs between the nebulizer, which is the ES probe, and the capillary. Aerosol of charged droplet formation at right angles to the capillary.

Ionization takes place as a result of imparting the strong electrical field and the superheated nitrogrn sheath gas, with Agilent AJS (Jet Stream) technology.

...

The influence of electrospray ion source design on matrix effects

March 17, 2022
Sensitivity, Assessment of Matrix Effects

Conclusion:
The Turbo Ion Spray (off-axis geometry), Turbo VTM Source (orthogonal geometry) and the Standard Z-Spray Source (double orthogonal geometry) did not differ much in their susceptibility to MEs.

The Jet Stream ESI (orthogonal geometry) reaches a higher sensitivity by an additional heated sheath gas, but suffers at the same time from significantly stronger signal suppressions than the comparable Standard ESI (orthogonal geometry) without sheath gas. No relation between source geometry and extent of signal suppression was found in this study.

...

Restek - Effective LC Troubleshooting Symptom-based Strategies and Solutions

January 12, 2022
Sensitivity

Summary of sensitivity reduction

  1. Simple potential causes:
    1.1) sample preparation,
     abnormal storage temperature,
     storage time,
     abnormal extraction,
     abnormal clean-up,
    1.2) LC system,
     Dilution error,
     autosampler needle not reaching sample,
     incorrect injection volume,
     abnormal LC back pressure,
     Leak,
     mobile phase unstable (modifier, prepared date, )
    1.3) Mass spectrometer system,
     non-asymmetric spray (due to clog)
     Electron multiplier voltage error,
     the declustering potential error,
     the collision energy error
     the collision exit cell potential error
     GS1 pressure - nebulizer gas error
     GS2 pressure - turbo gas error
     Curtain gas pressure error
     Ion spray voltage error
     test
    1.4) System monitoring,
      monitoring charts of internal standards, blank, calibrators,
    1.5) Data processing,
     Calculation error (dilution factor),
     Peak integration error (baseline issue, noise),
     Retention time shift,
         

  2. Complex potential causes:
     Analyze new known standard
     Analyze blank
     Infuse standard in 50-50 mobile phase solution to the mass spectrometer
     Dry run
         

...