Method Validation Workflow #
-
pre-validation testing
It could also be called method development -
Validation
determine accuracy, imprecision, sensitivity, linearity, interferences (specificity), matrix effects, and carryover
Order of Experiments #
- linearity, within-run imprecision, blank matrix, LLOQ
- matrix effect, recovery, interferences
- accuracy: certified reference materials, human samples, proficiency testing samples
- stability studies
- every run should include between-run imprecision samples
- carryover, dilutions
Accuracy #
evaluation of agreement between LC-MS/MS test results and true value.
Guidelines: FDA guidelines, CLSI EP9, CLSI EP15, CLSI C62
Imprecision #
evaluation of agreement between repeated analysis of same sample.
within-run imprecision
between-run imprecision
Sensitivity #
- depicted by lower limit of quantitation (LLOQ) of a method
- lowest concentration that can be reproducibly measured
- determined by the imprecision: < 20% between-run imprecision at LLOQ (at minimum use 20 matrix-based replicates over 20 days)
- determined by signal-to-noise ratio: >20:1 at LLOQ is recommended; 10:1 minimum
- ion ratios still have to pass
Linearity #
ability of LC-MS/MS method to provide counts per second values that are directly proportional to concentration of analyte
should carry out linearity experiment for every matrix that will be acceptable for a given method
do not use serial dilutions to create material for linearity experiments