Starts with setting up goals. #
null
Elements of method development #
- Analyte.
- Is there isobaric compounds?
- Analyte pKa value.
- SPE
- sorbent.
- washing reagent, eluting reagent
- Instrument.
- LC column.
- condition of instrument.
Elements of method validation #
- Linearity (linear range)
- Verify with external QC
- Analyte recoveries at low end and high end
- Precision (CV)
Case study #
Method Goals
- Clean extract
- Without the need for derivatization
- Resolve isomeric/isobaric species
Elements of method development
- Analyte.
Methylmalonic acid (MMA)
Isobaric compound: succinic acid
MMA pKa
- SPE
- To select a type of sorbent (Luna® Omega PS C18) for acidic compounds
- Sorbent pKa
- Equilibrating reagent: 1mL of 1% acetic acid in water;
- Washing reagent: 0.5mL of 50:50 methanol/Water;
- Eluting reagent: 2 x 0.6mL 2% ammonium hydroxide in Methanol;
- Reconstitute with 200 uL of 0.1% formic acid in water (mobile phase A)
Table 1: SPE procedure
- Experimental conditions
Table 2: LC Conditions
Elements of method validation
same as mentioned above
phenomenex - Method Development for Underivatized MMA from Serum